I got a new Owon SDS7102 oscilloscope and I'm using the included oscilloscope probes with it.

The manual says that the proper calibration procedure is to let the oscilloscope work at constant outside temperature for 25 minutes, do the oscilloscope's internal calibration routine and then set the probe capacitance using the included tool. The manual also says that the correct capacitance should be checked often and corrected if necessary. What the manual didn't mention is how often often is.

I expect that checking probes before each important measurement would be needed, but I'm looking for some "rule of thumb" for general use. From my current experience, several hours are enough to see small loss of sharpness of the calibration signal.

Best Answer

The probe capacitance adjustment is used to match the probe time constant, which consists of the probe capacitance and the probe resistance (usually 9 Megohms), to the time constant of the oscilloscope vertical amplifier input (which is usually 1 Megohm and 10 to 20 picofarads). When the time constants match, the overall attenuation of the probe/oscilloscope will be 10 to 1 over a wide frequency range. Normal practice in most labs is to adjust probes when they are first used both to ensure that they and the oscilloscope are operating OK and that nobody made any adjustments inbetween. That adjustment should hold for a day or more unless the temperature changes drastically or the probe and/or oscilloscope are defective. Defective could mean that the variable capacitor in the probe is not mechanically stable enough to hold its setting.